Pneumatic 6 GHz TEM Cell
TC-5063C
TC-5063C, Pneumatic 6 GHz TEM Cell generates the Electro-Magnetic field for testing small RF devices such as wireless communication receiver, Mobile phone, etc. An external test signal applied through the input port of the TC-5063C generates a consistent and predictable TEM test field inside the cell. The radiation field from a device transmitting in the Cell can also be detected through the port using a test receiver. The unique compact and economical design is optimized for medium accuracy measurements beyond the standard TEM Cell frequency range.
Introduction

TC-5063C, Pneumatic 6 GHz TEM Cell generates the Electro-Magnetic field for testing small RF devices such as wireless communication receiver, Mobile phone, etc. An external test signal applied through the input port of the TC-5063C generates a consistent and predictable TEM test field inside the cell. The radiation field from a device transmitting in the Cell can also be detected through the port using a test receiver. The unique compact and economical design is optimized for medium accuracy measurements beyond the standard TEM Cell frequency range.


Theory of operation


TC-5063C 6 GHz TEM cell is made to work beyond the typical TEM Cell operating frequency range limited by cell resonance. A typical TEM Cell is a 2-port symmetrical device; RF voltage is applied to one port while the other port is terminated in 50 ohm while maintaining 50 ohm characteristic impedance along the cell. Due to expansion and contraction parts of the cell, the wave propagation beyond certain frequency is no more propagated by TEM mode alone and creates resonance. To eliminate the resonance problem, the half of the cell is replaced by the wave absorbing material. One commercial implementationis G-TEM cell. The size of the G-TEM design is too large for typical small device applications due to the type of absorber used. TESCOM borrowed the concept of G-TEM, but changed the termination implementation scheme, and designed a very compact broad band TEM Cell that can be used on a desktop.


The operation principle of TC-5063C is essentially the same as TEM Cell. The E-H field inside the test volume is proportional to the input voltage and inversely proportional to the cell height. If a radiating object is inserted inside the cell, the radiated wave toward input port is guided by the transmission line and picked up at the input with a receiver such as a spectrum analyzer. With this method, the RFI from a radiating Device can be measured quantitatively. Since this apparatus is very broadband, it has many applications in the area of EMI, EMS, receiver sensitivity test, etc.


Key Features

◆ Radiation and susceptibility test
◆ Broadband TEM Cell up to 6 GHz
◆ Small size, Small footprint for desktop application
◆ High effective shielding
◆ Specifically designed for all types of mobile phones
◆ Pneumatic Open / Close Construction
◆ RS-232C Open / Close control


 Applications  


▷ Receiver sensitivity testing, Transmitter radiated power testing
EMI and EMS tests for small 6 GHz devices





Specifications

General Specification


VSWR< 1.7 , 100 MHz ~ 6 GHz

Path Loss: 25 dB Typical @1.8 GHz dipole
Effective Cell Height: 220 mm
Field Strength at Center of Cell: 13 dB mV/meter at 1 mV input
RF Connectors without module: 1 N(f) topside, 1 SMA(f) outside and SMA(f) inside

Remote control: RS-232C, 3 wire, DB9(p)

Line voltage: 100 - 240 VAC, 50/60 Hz, 15 watt max

Main air connector:  6 mm OD hose, one-touch push-on fitting
Input air pressure:  5 to 10 bar
Dimension:    Inside 240(W) x 205(D) mm

Outside 344(W) x 420(D) x 725(H) mm, door closed. 614(D) mm, door open

Door Size: 176(W) x 130(H) mm
Weight: approx. 27 kg
Packing Size:  450(W) x 540(D) x 840(H) mm
Packing Weight:  Approx. 33 kg


Typical RF Shielding


100 to 2000 MHz:  > 80 dB
2000 to 3000 MHz:  > 70 dB
3000 to 6000 MHz:  > 60 dB


* The shield effectiveness below is measured when the blank panel is mounted; other I/O interface panel results a different shielding effectiveness of the TEM Cell.


Ordering Information

TC-5063C 6 GHz TEM Cell (including accessories below)

- Test Report
- RF Cable, SS-402, N(m) to N(m) 2 m (< 6 GHz), 1 pc

- DATA Cable, RS232C, DB 9(s)-DB 9(s) 2 m, 1 pc
- Power Cable, 1 pc
- Hand Valve, 1 pc
- Remote Switch, 1 pc


Optional Accessories

4011-0001 RF Cable, SS-402, N(m) to N(m) 1 m (< 6 GHz)
4011-0019 RF Cable, SS-402, N(m) to N(m) 2 m (< 6 GHz)
4011-0020 RF Cable, SS-402, N(m) to SMA(m) 2 m (< 6 GHz)
4003-0001 DATA Cable, RS232C, DB 9(s)-DB 9(s) 2 m

4008-0017 USB Cable, USB A(p) to USB A(p) 1 m
4008-0018 USB Cable, USB A(p) to USB A(s) 50 cm


Download
[Data Sheet]
TC-5063C Data Sheet
[Manual]
TC-5063C Pneumatic TEM Cell Operating Manual